Test Method - Keysight N1913B EPM Series Service Manual

Power meters
Table of Contents

Advertisement

2
Performance Tests
Figure 2-5
The procedure details the key presses required on the Keysight N1913B Product
NOTE
name sentence style. For the Keysight N1914B Product name sentence style, the
equivalent key presses should be performed on both channels.

Test Method

1 Connect the equipment as shown in
2 Unplug the power cord from the range calibrator. Eliminate ground loops to
56
11683A range
calibrator front panel
N1913B/N1914B
Product name
3458A digital
multimeter
Absolute accuracy test setup
enable the 11683A (Option H01) range calibrator to operate properly.
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
11683A range
calibrator rear panel
BNC T-joint connector
BNC cable
BNC cable
Banana
connector
Banana
connector
Figure
2-5.
33250A function
generator