Linearity (Average Path); Description; Equipment; Test Method - Keysight N1913B EPM Series Service Manual

Power meters
Table of Contents

Advertisement

Linearity (Average Path)

Description

Linearity over the full input voltage range of the measurement path is warranted
to provide assurance of instrument accuracy. This test measures linearity by using
a calibration DAC and a calibration ADC (built into the DUT) to stimulate and
compare performance of the average path against the measurement ADC,
returning the worst case percentage error.

Equipment

No test equipment required

Test method

1 Execute the internal linearity measurement procedure for channel A.
2 Read back the result of the measurement from the DUT.
3 If the DUT model number is N1914B, then repeat this procedure for channel B.
4 The test will take a few minutes to complete.
5 The measurement result should be less than 0.5 and greater than –0.5. The
This test can be performed manually via the commands:
NOTE
SERV:BIST:CW[1|2]:LIN
SERV:BIST:CW[1|2]:LIN:PERR?
(Refer to the N1913B/N1914B EPM Series Power Meters Programming Guide for
further details on the use of these commands)
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
optimum measurement result for this test is 0.
Performance Tests
2
59

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914b epm series

Table of Contents