Power On Test / Self Test Failure Troubleshooting - Keysight E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Troubleshooting
Troubleshooting Using Service Function

Power On Test / Self Test Failure Troubleshooting

Table 4-2
relationships of failed tests to probable faulty board assemblies. If power on
test failed is displayed or self test fail is displayed, replace the faulty board
assembly as shown in
Table 4-2
Power on test / self test failure troubleshooting information
Test
Test
Test Name
group
No.
CPU
1
CPU Test
2
CPU Test
System
1
A1 FPGA check
2
A2 FPGA check
3
A1 Revision check
4
A2 Revision check
5
A3 Revision check
Data
1
System Data
2
SysCal Data
3
Flash File system
check
4
Test Limit Data
Keysight E4980A/E4980AL Precision LCR Meter
represents the contents of the power on tests or self test and the
Table
4-2.
Error
Probable faulty board assembly
Code /
Message
A1
1 ~ ffff
1 ~ ff
0x01
###
0x02
##
0x04
###
0x08
##
0x10
##
0x01
###
0x02
##
0x04
0x08
A2
A3
Bias
Handler
IF
###
###
###
###
Scanner
PPMC
###
###
#
#
#
#
#
###
###
43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4980al