6
Repair Guide
Instructions
Step 6
Fit the display interface board onto the plastic mounting lugs
on the display support molding.
Step 7
Attach the inverter board to the display interface board using
the two screws removed earlier.
Step 8
Connect the inverter board to the display interface board using
the backlight cable assembly.
Note:
The cable must be tucked under the plastic clip to prevent any
fouling.
Step 9
Attach the calibrator assembly to the display support molding
using the three screws removed earlier.
Step 10
Carefully spread the EMI fingers outwards, ensuring they
extend beyond the edges of the hole in which the calibrator
assembly is fitted.
134