Keysight N1913B EPM Series Service Manual page 133

Power meters
Table of Contents

Advertisement

Repair Guide
6
Instructions
Visual
Step 5
Install calibrator assembly onto the display support, tighten 3
screws by using bit T6 torque driver with 3.0 lbs-in.
Front USB flex cable and earth cable assembly
Place the USB Assembly Front (N1913-68305) at position
shown. Install screw (0515-5114) and tighten using Bit T6
torque driver with 3 lbf.in
Place the Assembly Bracket at the position shown.
Sequence for USB Assembly Front Earth Cable Installation:
1) Earth Cable N1912-61018
2) Flat Washer 3050-2633
3) Screw 0515-5114
Tighten using Bit T6 torque driver with 3 lbf.in
Ensure the cable lug is touching the solder lead as shown.
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
133

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914b epm series