Installation

Analog input board information

If an INET loop is to be used, install the loop cables. If an INET is not
being used, leave the INET cable between an in and out in place.
Connect the appropriate HP•IBor RS•232cable to the connector board.
Replace all of the instrument's panels and covers.
Analog input board information
Input Voltage
Input Noise
Input Common Mode
Range
Common Mode Rejection
Ratio
Scaling

Post-install checklist

U
U
U
U
Refer to the HP 5890 SERIES II Operating Manual for column
installation and operating information.
52
Are the instrument supply gasses connected?
Have all the supply gas fittings been leak tested?
Have the signal output cables been connected?
Has the instrument power cord been connected?
-10mV to + 1V between + and - inputs
¿
10
V peak to peak
+/- 5V DC (measured + or - to Gnd)
80 db
1 HP 5890 display count + 16.2
At Range 6
At Range 6
1 mV input to board will produce a 1
1 mV input to board will produce a 1
mV output
¿