Controls
6.2

Control/illumination unit

1 Display
2 Power switch
3 On/off switch for illumination
4
4 Toggle switch for illuminator 1/2
18
1
2
3
Leica M525 F20/CT20 / Ref. 10 715 156 / Version 01
Terminals under the control/illumination unit
5
6
13
12
5 Socket for potential equalization
For connecting the Leica M525 F20/CT20 to a potential
equalization device. This is part of the customer's building
installation.
Observe the requirements of EN 60601-1 (§ 8.6.7).
6 Power supply connection
7 Fuse holders
8 Auxiliary power outlet (max. output power 100 VA)
For requirements of use, see technical data, chapter 13.2.
CAUTION
Connecting unauthorized secondary devices to the auxiliary
power socket can lead to damage to the Leica M525 F20/CT20
surgical microscope and to the secondary device!
Never connect secondary devices to the auxiliary power
X
X
socket unless they conform to the specifications.
Requirements of use, see chapter 13.2.
9 12 V terminal (DC power supply for camera)
10 RS 232 interface (MDRS and Leica service)
11 USB interface (Leica service)
12 CAN-Bus interface
13 Footswitch/handswitch
Only footswitches and handswitches delivered by Leica
Microsystems (Schweiz) AG, Medical Division, may be connected
here.
Only systems verified by Leica Microsystems (Schweiz) AG,
Medical Division, may be connected to the connections
mentioned above.
7
8
11
10
7
9