Symbols

Class I Equipment (IEC 60601-1)

Accessible conductive parts cannot become live in the event of a basic insulation failure
because of the way in which they are connected to the protective earth conductor.

Type CF Equipment (IEC 60601-1)/Defibrillator Proof

This generator provides a high degree of protection against electric shock,
particularly regarding allowable leakage currents. It is type CF isolated
(floating) output and may be used for procedures involving the heart.
This generator complies with the IEC 60601-1:1988 +A1:1991 + A2:1995 and IEC 60601-
1:2005 specifications for "defibrillator proof" designation and IEC 60601-2-2:2006 and
IEC 60601-2-2:2009.
Liquid Spillage (IEC 60601-2-2:2006 Clause 44.3 and

IEC 60601-2-2:2009 Clause 201.11.6.3)

The generator is constructed so that liquid spillage in normal use does not wet electrical
insulation or other components which when wetted are likely to adversely affect the safety
of the equipment.
Voltage Transients (Emergency Generator Mains Transfer)
The generator continues to operate normally with no errors or system failures when
transfer is made between line AC and an emergency system-voltage source. (IEC 60601-
1:1988 + A1:1991 + A2:1995 clause 49, IEC 60601-1:2005 clause 11.8, IEC 60601-2-
2:2006 clause 51.101, and IEC 60601-2-2:2009 clause 201.11.8)
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator