Required Test Equipment; Service Screen Access; Figure 7-1 Key Code Screen - Covidien WarmTouch Service Manual

Convective warming unit
Table of Contents

Advertisement

Performance Verification

Required Test Equipment

7.3
Temperature probe (either handheld meter or USB-based):
Electrical leakage tester
Ground bond tester

Service Screen Access

7.4
The warming unit provides screens for performing functions restricted to
service personnel. These screens are protected by a key code.
When you see the message "Enter Key Code" press the keys indicated in
Figure 7-1
7-2
Accuracy of probe must be ±0.25°C or better over the range 10°C to 60°C.
Probe cable must be long enough to reach from computer to hose duct
adapter screen inside compressed hose.
simultaneously to access these screens.
Figure 7-1.  Key Code Screen
Service Manual