Performance Verification
Required Test Equipment
7.3
Temperature probe (either handheld meter or USB-based):
•
–
–
Electrical leakage tester
•
Ground bond tester
•
Service Screen Access
7.4
The warming unit provides screens for performing functions restricted to
service personnel. These screens are protected by a key code.
When you see the message "Enter Key Code" press the keys indicated in
Figure 7-1
7-2
Accuracy of probe must be ±0.25°C or better over the range 10°C to 60°C.
Probe cable must be long enough to reach from computer to hose duct
adapter screen inside compressed hose.
simultaneously to access these screens.
Figure 7-1. Key Code Screen
Service Manual