Class I Equipment (Iec 60601-1); Type Cf Equipment (Iec 60601-1)/Defibrillator Proof; Liquid Spillage - Covidien Force FX Service Manual

Electrosurgical generator c
Table of Contents

Advertisement

Standards and IEC Classifications
Symbol
Consult
instructions
for use

Class I Equipment (IEC 60601-1)

Accessible conductive parts cannot become live in the event of a basic insulation failure
because of the way in which they are connected to the protective earth conductor.

Type CF Equipment (IEC 60601-1)/Defibrillator Proof

Liquid Spillage

The Force FX Electrosurgical Generator C enclosure is constructed so that liquid spillage in
normal use does not wet electrical insulation or other components which when wetted
are likely to adversely affect the safety of the equipment.
3-16
Definition
Serial number
Indicates a serial number so that a specific medical device
can be identified
Authorized representative in the European community
Indicates the authorized representative in the European
community
Consult instructions for use
Indicates the operating instructions should be considered
when operating the device
The Force FX Electrosurgical Generator C provides a high
degree of protection against electric shock, particularly
regarding allowable leakage currents.It is type CF isolated
(floating) output and may be used for procedures involving
the heart.
The Force FX Electrosurgical Generator C patient return
electrode terminal is protected from defibrillator discharge
according to ANSI/AAMI HF18 and IEC 60601-2-2.
Reference
ISO 7000-2498
ISO 15223 5.1.2
ISO 7000-1641
Force FX Electrosurgical Generator C Service Manual

Advertisement

Table of Contents
loading

Table of Contents