Table 5-1 Ds-7000 Read Commands - Teledyne BENTHOS UDB-9000 User Manual

Universal deck box
Table of Contents

Advertisement

COMMAND NAME
Receive Pulse Duration
Receive Sensitivity
Version Information
Date
Time
Time Stamp
Transmit Frequency
Display Lockout
Power Output
Repeat Interval
Repeat Ping Status
Silent Ping
Ping Duration
Ping Lockout
First Reply
Turnaround Compensation
Frequency Filter Status
Frequency Filter
Reply
Repetitive Ping Header

Table 5-1 DS-7000 Read Commands

MNEMONIC
RD
RD=#.## where x is 6.25 or 12.50 in ms
RS
RS=### where ### is 010 to 256 in standard deviations over average noise level
VE
VE=s.s f.f where s.s is the display software version and f.f is the modem firmware ver-
sion
DA
DA=mm/dd/yy
TI
TI=hh:mm:ss
TS
TS=mm/dd/yy hh:mm:ss
TX
TX=##.## where ##.## is 07.00 to 15.00 in .25 increments
DL
DL=c where c is E for enabled or D for disabled
PO
PO=# where # is 0 through 8
RI
RI=##.# where ##.# is 02.0 through 99.0
RP
RP=c where c is E for enabled or D for disabled
SP
SP=c where c is E for enabled or D for disabled
PD
PD=##.# where ##.# is 10.0 to 25.0
PL
PL=### where ### is 000 to 995 in steps of 5
FR
FR=c where c is E for enabled or D for disabled
TC
TC=##.# where ##.# is 00.0 to 99.9 ms
FS
FS=c where c is E for enabled or D for disabled
FF
FF=##.## [##.##...] or FF=NONE where ##.## is 07.00 to 15.00 in .25 increments
RT
RT=## rx.ff xxx.xxxx where ## is the ping identifier, rx.ff is the receive frequency, and
xxx.xxxx is the reply time. ## will be 00 if no identifier was specified.
PR
PR=tx.ff c mm/dd/yy hh:mm:ss where tx.ff is the transmit frequency, and c is S for
silent ping or N for normal ping.
RESPONSE