SENSe:DISPlay

Description
Syntax
Query Syntax
Parameter/
Return parameter
Example

TRIGger:SOURce

Description
Syntax
Query Syntax
Parameter/
Return parameter
Example

READ

Description
Query Syntax
Return parameter
Example
Note
Sets or returns the display mode. There are two display
modes, normal and simple.
SENSe:DISPlay <NR1> | {OFF|ON}
SENSe:DISPlay?
<NR1>
OFF
ON
SENS:DISP OFF
Sets the display mode to normal.
Sets or returns current trigger source.
TRIGger:SOURce {INT|EXT}
TRIGger:SOURce?
INT
EXT
TRIG:SOUR EXT
Sets the current trigger source to external trigger.
Returns the measurement value.
READ?
<NR3>
READ?
>+2.2012E+0
Returns the measurement.
+9.0000E+9: it indicates measure value is OverRange.
+9.9999E+9: it indicates the HVP is detected.
COMMAND OVERVIEW
0:OFF.
1:ON.
Display mode is normal.
Display mode is simple.
Internal trigger mode.
External trigger mode.
± 0.0000~5.1000E± X
Set
Query
Set
Query
Query
101