7.6.3 Heater Door Spool Assembly

Refer to
"Figure 7-39 Heater Door Cable
Keeper plate
Cams
Button head screw
Spool assembly
To access the heater door spool assembly first disassemble the canopy/heater from the device as
described in "Removing the Canopy and Heater Assembly."
1.
Remove the housing bracket from the heater assembly.
2.
To replace the spool assembly, use a 2.5 mm hex key to remove the button head screw that secures the
cams and door links. Then use an 8 mm wrench to remove the 2 nuts that hold the spool.
Service Manual
Adjustment".
Right (East) Side
Part positions shown are from the back of the device when the doors are closed.

FIGURE 7-39. Heater Door Cable Adjustment

Radiant Heater Assembly Repair Procedures
Nylon washer (gray)
Inner cam (white)
Radiant heater arm
Outer cam (white)
Apply Lubriplate here.
Flat washer
Split ring lock washer
M4 screw
Heater door
opening rod
Bracket
Nuts
Chapter 7: Repair Procedures
Left (West) Side
Keeper plate
Top nut and flat washer
Flat washer
Lock washer
Bottom nut
7-39