9.16.2 Replace
bellows base latch
assembly
Ml110140
08/07
9 Repair Procedures
To replace the latch assembly, you must disassemble the bellows base assembly
to 由e
point where you can remove the guide (A) and latch assembly (8) as a unit
1.
Remove 伽e
BaglVent switch cartridge (Section 9.16.1).
2. Remove the two remainingscrews (C)that hold the AP L/ B 1V manifold to the bellows
base assembly. Remove the AP L/ B 1V manifol.
c
3. To remove the guide/latch assembly , remove two mounting screws (D) from the
underside. Remove two additional mounting screws from the topside. Remove the
guide/latch assembly from the bellows base assembly.
4. Separate the latch assembly from the guide assembly.
5. To install the new latch
a臼embly,
putthe spring
(日 inωplace
in the guide assembly
(l ong leg down).
6. Place the latch assembly on the guide assembly so
that 由 e
latch engages the short
leg of the spring. Secure the latch assembly
(町 to
the guide assembly.
E
F
7. Mount the guide/latch assembly into the bellows base assembly.
8. Reassemble the
brea由 ingsystem
in reverse order.
9. Perform the checkout procedure (Section 3).
9.45