Background
Interface and pin
assignment
Pin Assignment
The Scan function is used to automatically bin groups of
up to 100 components. The associated pins in the handler
interface are active when the Scan function is activated.
There are a total of 6 outputs, 3 inputs as well as a GND
and power (+5V) pin.
25Pin D-SHELL
(Female)
Controls the relay output.
Relay
Pass signal. Indicates the compare result
Pass
is IN(pass).
Low signal. Indicates a LO compare
Low
result.
High signal. Indicates a HI compare
High
result.
The clock signal will pulse high when
Clock
each group of output signals (Relay,
Pass, Low, High) are ready. There are up
to 100 groups of output signals
After all (100) output groups are ready,
STRB
the STRB signal will pulse high.
HANDLER/SCAN INTERFACE