System:pwm:on; System:pwm:off - Teledyne T3MIL50 User Manual

D.c. milli-ohm meters
Table of Contents

Advertisement

Example

SYSTem:PWM:ON

Description
Note
Syntax
Query Syntax
Parameter/
Return parameter
Example

SYSTem:PWM:OFF

Description
Syntax
Query Syntax
Parameter/
Return parameter
Example
126
SYST:LFR 60
Sets the line frequency to 60Hz.
SYST:LFR?
>60Hz
Returns the line frequency as 60Hz.
Sets or returns the duty ON period for the PWM drive
mode.
PWM drive mode is only available for the T3MIL50X.
SYSTem:PWM:ON <NR1>
SYSTem:PWM:ON?
<NR1>
SYST:PWM:ON 5
Sets the duty ON time to 5 adc units.
Sets or returns the duty OFF period for the PWM drive
mode.
SYSTem:PWM:OFF <NR1>
SYSTem:PWM:OFF?
<NR1>
SYST:PWM:OFF 200
Sets the duty OFF period to 200 ms.
T3MIL Series User Manual
3~99
Unit: time units. For 60Hz LF, each
unit is equal 16.6ms. For 50Hz LF,
each unit is equal to 20.0ms.
100~9999
Unit:ms
Set