Measure Delay

Background
1. Select Measure
Delay setting
2. Measure Delay
setting appears
Note
The Measure Delay setting inserts a delay time between
each measurement. Measure delay is turned off by
default.
Measurement start with
Measure delay time
Default Measurement
start time
The measure delay setting is useful for measuring
components that need some time to charge if the default
measurement start time is not adequate. An adequate
delay time allows the meter to avoid the effects of
transient disturbances that are usually seen when
measuring reactive DUTs with a current source.
From one of the main screens, press
the
key so that the menu
ESC
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter.
Go to Measure Delay and press
Enter.
Use the arrow keys to turn Measure Delay on and set the
delay time. Press Enter to confirm the setting.
Measure delay
setting
Measure Delay* OFF, ON: 000.000 ~ 100.000s
* When the set value is > 0.1s, the resolution is 0.1s.
When the set value is < 0.1S, the resolution is 1mS.
Pressing ESC before pressing ENTER will exit the
Measure Delay settings.
MEASUREMENT
Test signal
Measure delay time
Meas. Setup
menu icon
Enter
Move
Select menu
or setting
57