Temperature:ambient:data; System:lfrequency - Teledyne T3MIL50 User Manual

D.c. milli-ohm meters
Table of Contents

Advertisement

Parameter/
Return parameter
Example

TEMPerature:AMBient:DATa

Description
Syntax
Query Syntax
Parameter
Return parameter
Example

SYSTem:LFRequency

Description
Syntax
Query Syntax
Parameter/
Return parameter
<NR1>
OFF
ON
TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.
Sets or returns the user-set ambient temperature value for
the temperature compensation and the temperature
conversion function.
TEMPerature:AMBient:DATa <NRf>
TEMPerature:AMBient:DATa?
<NRf>
<NR2>
TEMP:AMB:DAT 25.6
Sets the user ambient temperature value to +25.6ºC.
TEMP:AMB:DAT?
>25.6
Returns the set ambient temperature as 25.6ºC.
Sets or returns the frequency setting for the line filter.
SYSTem:LFRequency {AUTO | 50 | 60}
SYSTem:LFRequency?
AUTO
50
60
COMMAND OVERVIEW
0:OFF.
1:ON.
Disables the user-set ambient
temperature.
Enables the user-set ambient
temperature.
-50.0~399.9 (Unit: ºC)
-50.0~399.9 (Unit: ºC)
The frequency setting for the line filter
is automatically detected.
The frequency is 50Hz.
The frequency is 60Hz.
Set
Query
Set
Query
125

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3mil50x

Table of Contents