FAQ

F
AQ
● What are the different measurement speeds?
● The T3MIL50/50X performance does not match the specifications.
What are the different measurement speeds?
There are two measurement speeds for both resistance and temperature
measurement. At the slow measurement rate, the measurement speed is 10
samples/s and at the fast measurement rate the measurement speed is at 60
samples/s.
The T3MIL50/50X performance does not match the specifications.
Make sure the device is powered on for at least 30 minutes, is operated at the
slow measurement rate and is within +18°C~+28°C with a humidity not
exceeding 80%. This is necessary to stabilize the unit to match the
specifications.