Binning:limit:reference; Binning:limit:result - Teledyne T3MIL50 User Manual

D.c. milli-ohm meters
Table of Contents

Advertisement

Parameter/
Return parameter
Example

BINNing:LIMit:REFerence

Description
Syntax
Query Syntax
Parameter
Return parameter
Example

BINNing:LIMit:RESult

Description
Query Syntax
Return parameter
Example
ABS
DPER
BINN:LIM:MODE DPER
Sets the mode to Δ%.
Sets or returns the limit reference value for the bin
sorting function.
BINNing:LIMit:REFerence {<NRf>[,<String>]}
BINNing:LIMit:REFerence?
<NRf>
<String>
<NR3>
BINN:LIM:REF 100
Sets the limit reference value to 100Ω.
BINN:LIM:REF?
>100.0000E+0
Returns the reference as 100Ω.
Returns the bin sorting function test result.
BINNing:LIMit:RESult?
<NR1>
BINN:LIM:RES?
>1
Indicates a pass for bin1.
COMMAND OVERVIEW
The test results are judged from
absolute values.
The test results are judged from a
reference value ± a percentage offset.
(delta percent)
000.0001~999.9999
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
000.0001~999.9999E± X
1~8: Bin1~Bin8
9: Bin Out
Set
Query
Query
111

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3mil50x

Table of Contents