Calculate:compare:percent:lower; Calculate:compare:percent:upper - Teledyne T3MIL50 User Manual

D.c. milli-ohm meters
Table of Contents

Advertisement

Syntax
Query Syntax
Parameter
Return parameter
Example
Note

CALCulate:COMPare:PERCent:LOWer

Description
Syntax
Query Syntax
Parameter
Return parameter
Example
Note

CALCulate:COMPare:PERCent:UPPer

Description
Syntax
Query Syntax
104
CALCulate:COMPare:LIMit:UPPer {<NRf>[,<String>]}
CALCulate:COMPare:LIMit:UPPer?
<NRf>
<String>
<NR3>
CALC:COMP:LIM:UPP 0.123,maohm
Sets the upper limit value to 0.123MΩ.
CALC:COMP:LIM:UPP?
>0.1230E+6
Returns the upper limit as 0.123MΩ.
This command is only applicable when compare mode is
set to ABS for compare function.
Sets or returns the lower limit percent value for the
compare function.
CALCulate:COMPare:PERCent:LOWer <NRf>
CALCulate:COMPare:PERCent:LOWer?
<NRf>
<NR2>
CALC:COMP:PERC:LOW 10.00
Sets the lower limit percent value to -10.00%.
CALC:COMP:PERC:LOW?
>10.00
Returns the lower limit as -10.00%.
This command is only applicable when compare mode is
set to DPER or PER for compare function.
Sets or returns the upper limit percent value for the
compare function.
CALCulate:COMPare:PERCent:UPPer <NRf>
CALCulate:COMPare:PERCent:UPPer?
T3MIL Series User Manual
000.0000~999.9999
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
000.0000~999.9999E± X
000.00~999.99
000.00~999.99
Set
Query
Set
Query