*The temperature coefficient for the other settings must be calculated
individually according to different conditions.
*If the temperature coefficient or the difference between the environmental
temperature and the required temperature exceeds normal operation, after
calculating the compensation, the variation to the reading value will be
significant.
*When using the PT-100 temperature sensor for temperature measurements,
the accuracy of the sensor (typical accuracy of <±0.5ºC) should also be taken
into account and calculated for.

Interface

Handler interface*
Scan*
Communication
Interfaces

Environmental

Operation
Environment
Storage Conditions

General

Power source
Accessories
Dimension
Weight
Signal: Trigger: TTL input
Signal: LOW, HIGH, FAIL, PASS, EOT, READY, BIN 1~8,
BIN OUT: total 15 TTL outputs.
Signal: RELAY, PASS, LOW, HIGH, CLOCK, STRB total 6
TTL outputs.
T3MIL50: USB/RS-232
T3MIL50X: USB/RS-232/GPIB
*The Scan and Handler interface use the same
connector.
Indoor use, altitude up to 2000m.
Operation Environment: 0˚C to 40˚C.
Temperature Range: 0 ~ 35˚C, Relative Humidity:
<80%RH; >35˚C, Relative Humidity: <70%RH.
Pollution Degree
-10˚C to 70˚C.
Temperature Range: 0 ~ 35˚C, Relative Humidity:
<90%RH; >35˚C, Relative Humidity: <80%RH
AC 100-240V±10%, 50-60Hz, 25VA
Power cord x 3
Test Lead: T3TL4K-150 x 1
223(W)×102(H)×283(D) mm
Approx. 3 kg
2
APPENDIX
147