Replacement Procedure

Outer Cover Replacement

Outer Cover Replacement

Tools Required

Removal Procedure

Refer to Figure 4-1 for this procedure.
Step 1. Disconnect the power cable from the E5061B.
Step 2. Remove the four bottom feet (item 1).
Step 3. Remove the two TORX T20 screws (item 2) fastening the side strap handles.
Step 4. Remove the four TORX T15 screws (item 3) fastening the four rear foot.
Step 5. Remove the four TORX T15 screws (item 4) fastening the cover.
Step 6. Slide up the outer cover and remove it carefully.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
NOTE
Set a torque to 14 in-lb for torque driver when fasten the TORX T15 screws (item 3 and
item 4).
54
torque screwdriver, T20
torque screwdriver, T15 (set to 14 in-lb)