Electromagnetic Immunity: Professional Healthcare Facility And Home Healthcare Environments - ZOLL M2 Operator's Manual

Table of Contents

Advertisement

Specifications
Electromagnetic Immunity: Professional Healthcare Facility and Home
Healthcare Environments
Functions of the ZOLL M2 include: ECG waveform monitoring from leads or pads, the pacing
pulse output, QRS detection, defibrillation energy discharge, and shock advisory functions.
Guidance and manufacturer's declaration – electromagnetic immunity –
for equipment and systems
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
a. The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
A-30
IEC 60601 test level
3 Vrms
150 kHz to 80 MHz
a
outside ISM bands
6 Vrms
150 kHz to 80 MHz
a
in ISM bands
10 V/m
80 MHz to 2.7 GHz
20 V/m (only for Defibrillation)
www.zoll.com
Compliance level
3 Vrms
6 Vrms
10 V/m
20 V/m
9650-000860-01 Rev. C