Emissions And Immunity Compliance - ZOLL 330 Service Manual

Multifunction aspirator
Hide thumbs Also See for 330:
Table of Contents

Advertisement

Specifications

Emissions and Immunity Compliance

The ZOLL Aspirator is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
The device design addresses EMC performance. Be sure to use the proper PWB identified in
this manual and make sure the proper power supply is supplied with this product.
The design does not rely on EMC gaskets or additional shielding to achieve compliance.
The operation manual identifies all applicable warnings for use in relation to Emissions and
Immunity Compliance.
Compliance for each emissions and immunity standard or test specified:
Emissions Test
The ZOLL aspirators are intended for use in the electromagnetic environment specified below. The
customer or user of the aspirator should ensure that they are used in such an environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEF 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Immunity Test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
B-4
Compliance
Electromagnetic Environment Guidance
Group 1
The ZOLL aspirators use RF energy only for their internal
function. Therefore, their RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B
The ZOLL aspirators are suitable for use in all
establishments other than domestic and those directly
Class B
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Complies
Test Level
+/- 8 kV contact
+/- 15 kV air
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
0% Ut, 0.5 cycles at 0, 45, 90,
135, 180, 225, 270, and 315
degrees; 0% Ut, 1 cycle and
70% Ut, 25/30 cycles; single
phase at 0 degrees
www.zoll.com
Compliance Level
+/- 8 kV contact
+/- 15 kV air
+/- 2 kV for power supply lines
+/- 1 kV for input/output lines
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
0% Ut, 0.5 cycles at 0, 45, 90,
135, 180, 225, 270, and 315
degrees; 0% Ut, 1 cycle and
70% Ut, 25/30 cycles; single
phase at 0 degrees
9650-000278-01 Rev. B