ZOLL M2 Operator's Manual page 208

Table of Contents

Advertisement

Messages
Alarm Message
BATTERY
COMMUNICATION FAULT
BATTERY ERROR
CARDIOVERSION USING
LEADS ECG - DISABLED
CHECK PADS -- PADS
SHORTED
CHECK PATIENT
CLEAR BUTTON STUCK
CODE MARKER BUTTON
STUCK
CONNECT THERAPY
CABLE
DEFIB DISABLED
DEFIB DISABLED -
SERVICE REQUIRED
DEFIB SERVICE
RECOMMENDED
DEMAND PACING
DISABLED
DIAG BUTTON STUCK
DISARM BUTTON STUCK
ECG ASYSTOLE
ECG C LEAD OFF
ECG COMMUNICATIONS
FAULT
ECG LEAD BUTTON
STUCK
C-2
Alarm Type
Priority
Equipment
Low
Equipment
Low
Equipment
Medium
Equipment
Low
Patient
High
Equipment
Medium
Equipment
Medium (Monitor/Pacer
mode;
High (Defib mode)
Equipment
Low
Equipment
High (Defib mode);
Low (Pacer/Monitor mode)
Equipment
High (Defib mode);
Low (Pacer/Monitor mode)
Equipment
Low
Equipment
Medium
Equipment
Medium (Monitor/Pacer
mode;
High (Defib mode)
Equipment
High
Patient
High
Equipment
Medium
Equipment
Medium
Equipment
Medium (Monitor/Pacer
mode;
High (Defib mode)
www.zoll.com
Cause
Battery communication fault.
A battery fault has been detected.
Lead ECG communication failed.
Pads shorted (when pads are selected
as primary ECG waveform in Monitor
mode).
VF/VT signal is detected in Monitor/
Manual Defib mode when the HR
alarm is on.
Front panel quick access key self-test
failed.
Front panel quick access key self-test
failed.
MFC is not connected to the unit (when
pads/paddles are selected as primary
ECG waveform in Monitor mode).
Cannot finish defibrillator charge in low
battery condition.
Defibrillator critical fault.
Total shock count exceeds 5000
shocks.
Power supply self-test failed, Pace/
Defib blank self-test failed.
Front panel quick access key self-test
failed.
Front panel quick access key self-test
failed.
Lead ECG HR detects asystole.
ECG C lead is not connected to the
patient when other leads are
connected.
There is an ECG communication fault.
Front panel quick access key self-test
failed.
9650-000860-01 Rev. C