Guidance And Manufacturer's Declaration - Electromagnetic Emissions - ZOLL AED PLUS Administrator's Manual

Defibrillator
Hide thumbs Also See for AED PLUS:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
Table 8: EMC Specifications
The ZOLL AED PLUS is intended for use in the electromagnetic environment specified below. The
customer or the user of the ZOLL AED PLUS should assure that it is used in such an
Emissions test
RF emissions CISPR11
RF emissions CISPR 11
Harmonic Emission
IEC 61000 3-2
Voltage Fluctuations/Flicker
Emission
IEC 61000 3-3
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to EMC information provided in this document.
The ZOLL AED PLUS is intended for use in the electromagnetic environment specified below. The
customer or the user of the ZOLL AED PLUS should assure that it is used in such an
Immunity test
IEC 60601 test
level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV differential
IEC 61000-4-5
mode
+/- 2 kV common
mode
Compliance
Group 1
Class B
Not applicable
Not applicable
Compliance
level
± 6 kV contact
± 8 kV air
Not applicable
± 1 kV I/O
Not applicable
Not applicable
ZOLL AED PLUS
environment.
Electromagnetic environment -
guidance
The ZOLL AED PLUS uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
environment.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.