Temperature
9650-0220-01 Rev. B

Advertisement

Table of Contents
loading

Summary of Contents for ZOLL M Series

  • Page 1 Temperature 9650-0220-01 Rev. B...
  • Page 2 The issue date or revision level for this operation guide is shown on the front cover. ZOLL and M Series are trademarks of ZOLL Medical Corporation.  2002 by ZOLL Medical Corporation. All rights reserved.
  • Page 3: Table Of Contents

    Enabling, Disabling and Suspending Alarms ....................4 ........................5 RINTING EMPERATURE ....................... 5 AINTAINING THE EMPERATURE PTION Performing the Daily Checkout Procedure ....................... 5 ............................6 PECIFICATIONS ............................7 OFTWARE ICENSE ..........................8 RDERING CCESSORIES ............................9 ROUBLESHOOTING ................................11 NDEX ZOLL M Series Temperature...
  • Page 4 (This page intentionally left blank.) ZOLL M Series Temperature...
  • Page 5: Preface

    Temperature option. Consult the Troubleshooting section if the M Series fails to operate as expected. If you do not know how to use the M Series, see the M Series Operator’s Guide and the relevant inserts. Thoroughly read the safety considerations and warnings sections in both the M Series Operator’s Guide and the relevant inserts before operating your M Series product.
  • Page 6: Indications For Use

    • Connect the ECG-out jack and modem (if available) only to other equipment with galvanically isolated circuits. • The cable and temperature probe are defibrillator proof when used as directed with the M Series. Using the Temperature option introduces no risk for shock due to defibrillation. Using the Temperature option introduces no risk for burns due to electro-surgery.
  • Page 7: Introduction

    2. M Series unit measures this resistance, converts it into degrees Celsius or Fahrenheit and continuously displays the temperature values. 3. If two temperature probes are connected, the M Series unit also calculates and displays the difference between the two measured temperatures ( T).
  • Page 8: Preparing To Take Measurements

    4. If you are using two temperature probes, connect Y-adapter to temperature connector on the front panel of the M Series unit (see Figure 1: on page 3). 5. Connect the temperature probe cable(s) to the temperature connector or Y-adapter.
  • Page 9: Reading The Display

    ∆ information: • Temperature • Alarm status • Unit of measurement NOTE Depending on the options included in your M Series, your front panel and display may slightly differ from the illustrations in this insert. Temperature Data Display Area Temperature...
  • Page 10: Setting Alarms

    T1 and T2 ( T) ∆ When the M Series is turned on, all alarm functions are disabled and the alarm limits are set to their default values. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off.
  • Page 11: Printing Temperature Data

    2. Set the simulator to simulate 68° F or 20° C. 3. Verify that the M Series displays 68° F (+/- 0.2° F) or 20° C (+/- 0.2° C). 4. Set the simulator to simulate 104° F or 40° C.
  • Page 12: Specifications

    Sampling Rate 1 sample/sec (1 Hz) Resolution 0.1° F (0.06° C) Table 4: Battery Operating Time Specifications (using ZOLL XL Battery Pack) For a new, fully charged battery pack at 60 defibrillator discharges at maximum energy (200J 20°C with IBP, EtCO...
  • Page 13: Software License

    Purchaser signifies agreement to and acceptance of the following terms and conditions: Grant of License: In consideration of payment of the software license fee which is part of the price paid for this product ZOLL Medical Corporation grants the Purchaser a non-exclusive license, without right to sublicense, to use the system software in object-code form only.
  • Page 14: Ordering Accessories

    Ordering Accessories To order accessories for the M Series Temperature option call ZOLL Customer Service at 1-800-348- 9011. Customers outside the United States should call the nearest authorized ZOLL Medical Corporation sales representative. The following tables list optional accessories that can be used with...
  • Page 15: Troubleshooting

    The following table lists the Temperature messages, the problem each message indicates, and the actions you should take to fix the problem. Read this section carefully before monitoring patients. For further assistance, call ZOLL Technical Service at 1-800-348-9011. Customers outside the United States should call the nearest authorized ZOLL Medical Corporation sales representative.
  • Page 16 (This page intentionally left blank.) ZOLL M Series Temperature...
  • Page 17: Index

    Temperature simulator Principle of operation 6 approved type 8 Printing Daily checkout procedure 5 ordering 8 temperature data 5 Temperature value, reading on display 4 Default types of measurements printed 5 alarm 4 Troubleshooting 9 Probe ZOLL M Series Temperature...
  • Page 18 Unit of measurement, reading on display 4 Warning anesthetics v cable routing v defibrillation v explosion v general v oxygen-rich atmosphere v Water, caution against immersion v Y-adapter See 2-channel Y-adapter ZOLL customer service phone number 8 technical service phone number 9 ZOLL M Series Temperature...