CHECK PULSE, IF NO PULSE PERFORM CPR
If configured to do so, this message is displayed and announced in the following situations:
At the start of the CPR interval
•
After a No Shock Advised analysis result
•
After a shock is delivered
•
If a shock is advised but not delivered
•
IF NO PULSE, PERFORM CPR
If configured to do so, this message is displayed and announced in the following situations:
At the start of the CPR interval
•
After a No Shock Advised analysis result
•
After a shock is delivered
•
If a shock is advised but not delivered
•
PERFORM CPR
If configured to do so, this message is displayed and announced in the following situations:
At the start of the CPR interval
•
After a No Shock Advised analysis result
•
After a shock is delivered
•
If a shock is advised but not delivered
•
STOP CPR
After performing CPR for the configured period, the unit will announce and display a STOP
CPR prompt immediately prior to restarting ECG analysis.
PUSH HARDER (Adult Patients only)
This message is announced when the chest compressions applied during CPR are not deep
enough compared to configured target depth settings (2.0" or greater in AHA/ERC
recommended protocols).
GOOD COMPRESSIONS (Adult Patients only)
This message is announced when the rescuer responds to a PUSH HARDER prompt by
increasing chest compression depth to or above the configured target depth.
CHECK PADS - PADS SHORTED
This message is displayed and CHECK PADS is announced when a pads shorted condition is
detected. This condition needs to be corrected before ECG analysis or defibrillation treatment
can be delivered. This message is displayed when the MFC is connected to the test plug used
for the 30 Joule self-test.
CHECK PATIENT
This message is displayed and announced when the unit detects a shockable rhythm after a full
Analysis/CPR cycle when the ZOLL M2 is configured to pause after each CPR period. In this
configuration, the ANALYZE button initiates the ECG analysis/CPR cycle.
9650-000860-01 Rev.C
ZOLL M2 Operator's Guide
8-11