Table B-7 Guidance And Manufacturer's Declaration - Electromagnetic Immunity For Platform - ZOLL AutoPulse 100 User Manual

Resuscitation system
Table of Contents

Advertisement

User Guide
Electromagnetic Immunity Declaration (EID)
Table B-7
Guidance and Manufacturer's declaration – Electromagnetic immunity for Platform
The Platform is intended for use in the electromagnetic environment specified below. The customer or user of the
Platform should ensure that it is used in such an environment.
Immunity test
Electrostatic Discharge
(ESD) IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field. IEC 61000-
4-8
IEC 60601 Test Level
± 8 kV contact
± 15 kV air
Not Applicable
Not Applicable
Not Applicable
30 A/m
Compliance Level
± 8 kV contact
± 15 kV air
Not Applicable
Not Applicable
Not Applicable
30 A/m
Electromagnetic
environment – guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%
No mains power connection
No mains power connection
No mains power connection
Power frequency magnetic
fields should be at levels
characteristic of typical
location in a typical
commercial or hospital
environment.