Chapter 11 External Pacing

3 Turn Mode Selector to PACER

Turn the Mode Selector to PACER. The Pacer Dashboard displays.
Note:
Navigating the Pacer Dashboard
When the unit is in Pacer mode, the Pacer Dashboard displays and the Output field is
highlighted. To navigate around the pacer dashboard, turn the trim knob to move to and
highlight a field. Press the trim knob to select the field, and turn the trim knob to change the
setting, then press the knob again to enter the selected setting.
If the pacer mode and rate settings are acceptable without making any changes, go to step 7 to
set the output current. Otherwise, go to step 4.

4 Set Pacer Mode

Use the Trim Knob to navigate to the Mode field in the Pacer Dashboard. Press the trim knob to
activate the mode field. Turn the trim knob to change the setting to Demand, and press the trim
knob to select it.

5 Set Pacer Rate

Use the Trim Knob to navigate to the Rate field in the Pacer Dashboard, press the knob, and
then turn the knob to set the Pacer Rate to a value 10-20 ppm higher than the patient's intrinsic
heart rate. If no intrinsic rate exists, use 100 ppm. You can increase or decrease the pacer rate in
increments of 2 ppm.
Note:
11-4
If the unit's default mode is AED mode and has been configured to require a password,
when the password screen displays, use the Trim knob to enter the password, and press
the knob to select OK.
The default pacing rate is 70 ppm. This default rate is configurable.