Chapter 13 Maintenance and Troubleshooting
Symptom
The device cannot switch
modes (non-AED use).
The ZOLL M2 unit cannot
print.
Defibrillation and Pacing Problems
Excessive Disarms Detected
30 J Test Failed
DEFIB DISABLED message
Button Stuck (Paddles button,
front panel Shock
button,Charge button,etc.).
DEMAND PACING
DISABLED message
PACER DISABLED message
PACER CURRENT OUT OF
TOLERANCE message
CHARGE FAILED message
13-12
Recommended Action
Turn the ZOLL M2 unit off and then turn it back on to
correct the fault. If the fault still exists, contact the
appropriate technical personnel or ZOLL Technical
Service Department.
• Close the printer door.
• Load paper into the printer.
• Verify orientation of paper in the drawer
• Take out the jammed paper and insert in proper
orientation.
• Stop using the printer, and wait until the printer
head/motor temperature drops to normal.
• Contact the appropriate technical personnel or
ZOLL Technical Service Department.
Wait longer between successive internal discharges
(disarm internally instead of to patient).
Turn the ZOLL M2 unit off and then on to correct the
fault. Rerun the 30 J test. If the fault still exists,
contact the appropriate technical personnel or ZOLL
Technical Service Dept.
Turn the ZOLL M2 unit off and then on to correct the
fault. If the fault still exists, contact the appropriate
technical personnel or ZOLL Technical Service
Department.
Press the stuck button, turn the ZOLL M2 unit off,
then turn it back on to correct the fault. If the fault still
exists, remove the device from use and contact the
appropriate technical personnel or ZOLL Technical
Service Department.
Turn the ZOLL M2 unit off and then on to correct the
fault. If the fault still exists, contact the appropriate
technical personnel or ZOLL Technical Service
Department.
Turn the ZOLL M2 unit off and then on to correct the
fault. If the fault still exists, contact the appropriate
technical personnel or ZOLL Technical Service Dept.
Continue to use Pacer function to treat patient. When
finished treating patient, turn the ZOLL M2 unit off and
then contact the appropriate technical personnel or
ZOLL Technical Service Department.
• Plug the device into AC power.
• Replace the battery pack with a fully charged
battery pack.
www.zoll.com
9650-000860-01 Rev. C