Chapter 5 Alarms
Alarm Indicator Self-Test
The ZOLL M2 unit performs a self-test of the audible and visual alarm indicators upon start up.
To ensure that the alarms are functioning properly, verify that you hear an alarm tone and the
indicator lights are illuminated for three seconds upon start up of the unit.
Patient Alarm Display
When a patient's vital signs measurements trigger an alarm, in addition to an alarm tone and the
illumination of indicator lights, the ZOLL M2 unit displays an alarm text message in the status/
alarm message field, and changes the display characteristics of the monitoring function's
numeric display (the alarming parameter appears in red text with white background).
Note:
In the following example, the heart rate (160 bpm) has risen above the upper alarm limit (HR
High Alarm):
Click + to
display more
alarms/
messages
Figure 5-1 Patient Alarm Display
.
5-2
When more than one alarm message occurs, the field to the left of the status/alarm
message field changes from "-" into "+". Click the plus sign to see the hidden alarm
messages.
Status / Alarm message field is red with white text
www.zoll.com
Alarming parameter displays
in red/white
9650-000860-01 Rev. C