Maintenance and Troubleshooting
Resuscitation equipment must be maintained to be ready for immediate use. To ensure the
readiness and optimum working condition of the ZOLL M2 unit, you should perform the
following inspections and tests daily or at each shift change.
In addition to the daily check, authorized personnel should complete performance and
calibration testing at regularly scheduled intervals, which should not exceed one year.
A maintenance log is an important part of a successful maintenance program in which you
record information on a regular basis. This allows for verification of necessary maintenance
and for scheduling periodic requirements such as calibration and certification.
Based on the recommendations of the Defibrillator Working Group of the U.S. Food and Drug
Administration
chapter (and can be copied for use as needed).
Warning!
Do not perform service on the ZOLL M2 unit while it is connected to a patient.
1. JAMA. 1990;264:1019-1025
9650-000860-01 Rev. C
1
, ZOLL suggests using an operator's shift check list, which is included in this
ZOLL M2 Operator's Guide
Chapter 13
13-1