configured to stay in Sync mode after defibrillation in the Basic Defib default settings in the
Supervisor System Main Menu.
The unit will not leave Sync mode when the following actions occur:
Allowing the Ready state to time out
•
Changing the selected energy levels
•
Pressing the Disarm key without pressing Shock
•
1 Select Energy Level
Press the ENERGY SELECT arrows up or down to select the desired energy level.These
buttons are located either on the front of the unit or on the STERNUM paddle.
WARNING!
When using pediatric defibrillation electrodes, the patient type must be set to Pediatric
and the defibrillator energies must be set based on site-specific institutional protocols
for pediatric defibrillation.
2 Charge Defibrillator
Press the CHARGE button on the front panel or on the APEX paddle handle.
To increase or decrease the selected energy after you have pressed the CHARGE button, use
the defibrillator ENERGY SELECT arrows on the front panel or sternum panel.
9650-000860-01 Rev.C
3
SHOCK
2
CHARGE
1
ENERGY
SELECT
3
SHOCK
2
CHARGE
1
ENERGY
SELECT
ZOLL M2 Operator's Guide
Synchronized Cardioversion Procedure
or
9-17