Electromagnetic Immunity (Iec 60601-1-2 - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Specifications

Electromagnetic Immunity (IEC 60601-1-2)

Guidance and manufacturer's declaration – electromagnetic immunity
The Propaq MD unit is intended for use in the electromagnetic environment specified below. The customer
or the user of the Propaq MD should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD) IEC
61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
T
A-30
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5% U
T
(>95% dip in U
T
0.5 cycle
40% U
T
(60% dip in U
) for
T
5 cycles
70% U
T
(30% dip in U
) for
T
25 cycles
<5% U
T
(>95% dip in U
T
5 sec
3 A/m
is the AC mains voltage prior to application of the test level.
www.zoll.com
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV common mode
<5% U
T
) for
(>95% dip in U
) for
T
0.5 cycle
40% U
T
(60% dip in U
) for
T
5 cycles
70% U
T
(30% dip in U
) for
T
25 cycles
<5% U
T
) for
(>95% dip in U
) for
T
5 sec
3 A/m
Electromagnetic environment
– guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Propaq MD unit requires
continued operation during
power mains interruptions, it is
recommended that the Propaq
MD unit be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
9650-0806-01 Rev. D