IBP option. Consult the Troubleshooting section if the M Series fails to operate as expected. If you do not know how to use the M Series, see the M Series Operator’s Guide and the relevant inserts. Thoroughly read the safety considerations and warnings sections in both the M Series Operator’s Guide and the relevant inserts before operating your M Series product.
Then check the catheter, tubing, transducer, cabling and M Series IBP option for proper functioning. • Do not immerse the M Series device, batteries, cables, or transducers in water, solvents, or cleaning solutions. • If the M Series device gets wet accidentally, wipe it down with a clean towel.
• Configure alarms to go off when the unit detects pressure values above or below user-programmable limits. How Do I Use the IBP Option? To take safe and accurate pressure measurements using the M Series IBP option, you must do all of the following: Select the proper transducer.
1. Inspect the transducer and the cable for wear, breakage or fraying. Replace cable or transducer if necessary. 2. Connect the transducer to the IBP interface cable, if applicable. 3. Connect the IBP cable to the IBP connector on the front panel of the M Series unit (see Fig. 1). Pressure Pressure...
Before you can zero the transducer, you must first display the IBP menu. You can display the IBP menu without leaving Monitor, Defib or Pacer mode. AEDs must be in manual mode to display the IBP menu. See the M Series Operator’s Guide for further details on activating the manual override mode.
Zeroing the Transducer To ensure that the M Series unit measures pressure accurately, you must zero the transducer before each use. The M Series automatically displays ZERO P1 or ZERO P2 when zeroing is necessary. If you change or disconnect a transducer, you must zero the new transducer before use. If you move a transducer to a different monitor, you must zero the transducer again even if you have already done so on another unit.
• Catheter clogging. • Air bubbles in catheter or transducer dome. Flush catheter regularly while taking IBP measurements. Always view the IBP AUTION trace to ensure pressure measurements are based on a physiological waveform. ZOLL M Series Invasive Blood Pressure...
To access most IBP features you must press a softkey with the proper softkey label (see Fig. 5). NOTE Depending on the options included in your M Series, your front panel and display may slightly differ from the illustrations in this insert.
Page 13
The M Series default unit of measure is millimeters of mercury (mmHg), but you can configure the M Series to display kilopascals (kPa). See the M Series CCT Configuration Guide for instructions on how to configure alternate units of measure.
Changing the Display You can change how the M Series unit displays IBP data and traces. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off.
4. Press the Select softkey to select the label you want to display. 5. Press the Enter softkey. M Series unit displays the selected label. If desired, repeat steps 1 - 5 to change the other label. ZOLL M Series Invasive Blood Pressure...
4. Press the Select softkey to select the range you want to display. 5. Press the Enter softkey. M Series unit displays the trace with the selected range. If desired, repeat steps 1 - 5 to change the other pressure channel range.
Setting Alar ms Before taking measurements, check that all IBP alarms settings are appropriate for the situation. You can set the alarms for Pressure channel 1, Pressure channel 2 and NIBP separately. See M Series Operator’s Guide for instructions on setting alarms.
-50 to 295 mmHg; -6.7 to 39.2 kPa When the M Series is turned on, all alarm functions are disabled and the alarm limits are set to their default values. Any changes to these settings remain in effect until either the settings are changed or 10 seconds after the M Series is turned off.
Printing IBP Data See the M Series Operator’s Guide Recorder Operation section for instructions on how to print data. If you are currently displaying IBP traces, press the RECORDER button to print a strip chart that includes the following values for the currently displayed trace: •...
Input: minimum 250 Ω Output: maximum 3000 Ω Transducer Impedances Transducer Excitation 2.5 V DC Table 7: Battery Operating Time Specifications (using ZOLL XL Battery Pack) For a new, fully charged battery 60 defibrillator discharges at maximum energy (200J Biphasic), pack at 20°C with IBP, EtCO or 1.75 hours minimum of continuous ECG, IBP (2 channels),...
Purchaser signifies agreement to and acceptance of the following terms and conditions: Grant of License: In consideration of payment of the software license fee which is part of the price paid for this product ZOLL Medical Corporation grants the Purchaser a non-exclusive license, without right to sublicense, to use the system software in object-code form only.
To order accessories for the M Series IBP option call ZOLL Customer Service at 1-800-348-9011. Customers outside the United States should call the nearest authorized ZOLL Medical Corporation sales representative. The following tables list optional accessories that can be used with the M Series IBP option:...
The following table lists the IBP messages, the problem each message indicates, and the actions you should take to fix the problem. Read this section carefully before monitoring patients. For further assistance, call ZOLL Technical Service at 1-800-348-9011. Customers outside the United States should call the nearest authorized ZOLL Medical Corporation sales representative.
Page 24
Reposition transducer and rezero as necessary. WARNING! * Air in the monitoring line risks accidental air embolization associated with significant patient morbidity. WARNING! ** Leaks in the tubing/catheter system may effectively cause the patient to hemorrhage. ZOLL M Series Invasive Blood Pressure...
Checkout procedure 13 Pediatric patients, indications vi Connecting transducer 2 Personnel, qualified for operation v Connector, location of 2 Phone number KPa, meaning of 7 Contraindications vi ZOLL customer service 16 CVP, meaning of 9 ZOLL M Series Invasive Blood Pressure...
Page 26
Trace display area description of 6 illustration 6 Trace Options menu, illustration 8 Traces menu, illustration 8 Transducer approved types 16 caution vi compatibility 14 connecting 2 excitation 14 impedance 14 indications vi selecting proper 2 ZOLL M Series Invasive Blood Pressure...