ZOLL M2 Operator's Manual page 184

Table of Contents

Advertisement

Specifications
Table A-2. Rated Delivered Energy at Every Defibrillator Setting into a Range of Loads
Selected
Energy
25
1 J
1 J
2 J
1 J
3 J
2 J
4 J
3J
5 J
3 J
6 J
4 J
7 J
5 J
8 J
5 J
9 J
6 J
10 J
7 J
15 J
10 J
20 J
14 J
30 J
21 J
50 J
35 J
70 J
49 J
85 J
60 J
100 J
71 J
120 J
85 J
150 J
107 J
200 J
142 J
The ZOLL M2 Rectilinear Biphasic waveform employs the same first and second phase timing,
the same first and second phase currents/voltages, and the same mechanisms for controlling
defibrillation waveshape as the ZOLL X Series
ZOLL M2 defibrillation waveforms are considered substantially equivalent.
Figures A-1 through A-20 show the Rectilinear Biphasic waveforms that are produced when
the ZOLL M2 defibrillator is discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms
at each energy setting.
A-14
Load
50
75
100
1 J
1 J
1 J
2 J
2 J
2 J
3 J
3 J
3 J
4 J
4 J
5 J
5 J
6 J
6 J
6 J
7 J
7 J
7 J
8 J
8 J
8 J
9 J
9 J
9 J
10 J
11 J
10 J
12 J
12 J
16 J
17 J
18 J
21 J
23 J
24 J
32 J
35 J
36 J
54 J
59 J
61 J
76 J
83 J
85 J
92 J
101 J
104 J
109 J
119 J
122 J
131 J
143 J
147 J
164 J
180 J
183 J
230 J
249 J
253 J
www.zoll.com
125
150
175
1 J
1 J
1 J
2 J
2 J
2 J
3 J
3 J
3 J
5 J
5 J
4 J
6 J
6 J
6 J
7 J
7 J
7 J
8 J
8 J
8 J
10 J
9 J
9 J
11 J
11 J
10 J
12 J
12 J
12 J
18 J
18 J
17 J
24 J
24 J
23 J
37 J
36 J
35 J
62 J
61 J
59 J
87 J
86 J
83 J
106 J
104 J
101 J
125 J
123 J
119 J
150 J
147 J
143 J
188 J
184 J
179 J
269 J
261 J
260 J
®
®
and R Series
defibrillators. The X Series and
9650-000860-01 Rev. C
Tolerance
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 3 J
+/- 15%
+/- 15%
+/- 15%
+/- 15%
+/- 15%
+/- 15%
+/- 15%
+/- 15%