Battery Status And Ac Power Indicators - ZOLL M2 Operator's Manual

Table of Contents

Advertisement

Chapter 2 Product Overview

Battery Status and AC Power Indicators

When the unit is plugged in to AC mains, the AC power LED illuminates.
The battery status indicator displays various battery icons to indicate the approximate
remaining unit run time based on the charge state of the battery. Additionally, these icons
provide status indications of the battery connection and communication with the unit. Not all
battery icons are shown in the table below; there are additional battery icons that display thirty
minute increments, such as 1:30+ and 2:30+.
Note:
Icon
2-6
Upon powering up the ZOLL M2 unit, the battery capacity will be displayed within a
short period under normal conditions. Under some circumstances, such as activating
the monitor/defibrillator immediately after the unit is turned on, the battery icon may
display less than one hour battery capacity for up to two minutes after exiting the
defibrillation mode.
Status
No battery detected
Communication fault
Battery Error
Low battery capacity
Battery Level 1
Battery Level 2
Battery Level 3
Battery Level 4
Battery Level 5
Battery calibration required
www.zoll.com
Indication/Action
Either there is no battery in the unit
while it is being powered by the AC
mains, or the device cannot detect
that the battery is connected. Install
new battery or replace existing
battery.
The unit is unable to establish
communication with the battery.
Check the battery contacts.
A battery fault has been detected.
Plug the ZOLL M2 unit into a power
source or install a new battery.
The battery has reached a low
battery state. Replace the battery
immediately.
The battery has enough energy to
operate the ZOLL M2 unit for more
than 30 minutes under current
operating conditions.
The battery has enough energy to
operate the ZOLL M2 unit for more
than one hour under current
operating conditions.
The battery has enough energy to
operate the ZOLL M2 unit for more
than two hours under current
operating conditions.
The battery has enough energy to
operate the ZOLL M2 unit for more
than three hours under current
operating conditions.
The battery is fully charged.
When the message, BATTERY
CALIBRATION REQUIRED
displays, the battery has reached a
state that requires battery
calibration. Calibrate battery using
the ZOLL SurePower Charger
Station.
9650-000860-01 Rev. C