Chapter 8 Automated External Defibrillator (AED) Operation

3 Press SHOCK

WARNING!
Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient's
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
Press and hold the illuminated SHOCK button on the front panel until energy is
delivered to the patient.
Observe the patient or ECG response to verify that the shock has been delivered.
The delivered energy level and the shock number (1) are displayed in the panel at the bottom of
the screen.

Perform CPR

Begin chest compressions and rescue breathing per local protocol. Follow the CPR metronome
by compressing the patient's chest in synchronization with the metronome beeps.
WARNING!
Place the patient on a hard surface before starting chest compressions. Compression
depth monitoring may not be accurate when compressions are performed with the
patient on a flexible surface such as a mattress.
Note:
8-8
If ZOLL CPR pads are connected and the patient is an adult, the unit monitors the rate
and depth of chest compressions and can display PUSH HARDER and GOOD
COMPRESSIONS messages and voice prompts.