3 Deliver Shock

WARNING!
Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient's
body to come into contact with metal objects, such as a bed frame, as unwanted
pathways for defibrillation current may result.
Apply each paddle to the patient's chest with a force of 10 - 12 kilograms (22 - 26.4 pounds) in
order to minimize patient impedance and achieve optimal results.
Note:
Use both thumbs to simultaneously press and hold both SHOCK buttons (one on each paddle)
until energy is delivered to the patient.
Note:
The delivered energy level and the shock number (1) are displayed at the bottom of the screen.
If additional countershocks are needed, follow steps 1 through 3 of this procedure starting on
page 9-3 to readjust the energy settings, charge the unit, and deliver the shock.
9650-000860-01 Rev.C
To cancel the defibrillation and disarm the defibrillator at any time, press the Disarm
quick access key (or press the Energy select up or down key).
The shock must be delivered within 15 to 60 seconds after the ready state begins
(depending on configuration), or the defibrillator will disarm itself.
The front panel SHOCK button (
Shock button(s). Pressing this button instead of the paddle SHOCK buttons produces
an audible invalid operation tone.
ZOLL M2 Operator's Guide
Emergency Defibrillation Procedure with Paddles
) is inactive when using external paddles with
9-7