Guidance and Manufacturer's Declaration — Electromagnetic Emissions

Electromagnetic Immunity Declaration (EID)

The ZOLL AED Pro device is intended for use in the electromagnetic environment specified
below. The customer or operator should ensure that the device is used in such an environment.
Immunity Test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
9650-0350-01 Rev. J
IEC 60601
Test Level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV for input/output
lines
±1 kV differential mode
±2 kV for common
mode
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 seconds
Note: Ut is the ac
mains voltage prior to
application of the test
level.
3 A/m
ZOLL AED Pro Operator's Guide
Compliance
Electromagnetic Environment –
Level
±6 kV contact
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
±8 kV air
the relative humidity should be at
least 30%.
Not applicable
±1 kV I/O
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Guidance
A–7