Page 5
ZOLL Medical Corporation’s M Series Service Manual is intended for the service technician who identifies malfunctions and makes repairs at the subassembly level. Place this insert into the three-ring binder containing the M Series Service Manual and all other M Series option inserts.
Page 6
:DUQLQJ sufficient current to cause lethal shocks. Do not discharge the M Series unit’s energy internally more than three times in one minute or damage to the unit may result. Do not discharge a battery pack except in a Base PowerCharger , a compatible ZOLL Battery Charging/Testing unit, or an M Series unit.
• 5HWXUQLQJ D XQLW IRU VHUYLFH If you need to send the unit to ZOLL Medical Corporation for repair, you must obtain a service request (SR) number from the Technical Service representative. Units are available on loan while your unit is being repaired.
ECG trace. The IBP/Temperature module resides in the lower housing assembly of the M Series CCT unit. A transducer connects to the M Series CCT unit through one of two pressure channel connectors on the front of the unit.
/22' 5(6685( When the M Series CCT unit detects a connection to a transducer through one or both of the pressure channel connectors, the unit prompts the user to zero the pressure channel(s). Pressure measurement does not begin until the pressure channels have been zeroed.
• 5HFRUGLQJ \RXU UHVXOWV Photocopy the checklist at the back of the M Series Service Manual and use the copy to record the results of the maintenance tests. As you complete each step in a procedure, mark the appropriate Pass/Fail/NA box, and keep the checklist for your maintenance file. Complete all the steps of the procedure before evaluating the test results.
Set the BP output sensitivity scale to 5 V/V/mmHg on the Patient Simulator. µ • Connect the Patient Simulator cables to the IBP connectors on the M Series CCT unit. • 3URFHGXUH Follow these steps to perform the IBP monitor test.
Set the BP output sensitivity scale to 5 V/V/mmHg on the Patient Simulator. µ • Connect the Patient Simulator cables to the IBP connectors on the M Series unit. • 3URFHGXUH Follow these steps to perform the IBP noise test.
Transducer Simulator , and set the • gain to x10. Connect the Transducer Simulator cable to the P1 connector on the M Series CCT unit. • 3URFHGXUH Follow these steps to perform the IBP static pressure test. Note: These instructions apply to the Fogg BP-28 Pressure Transducer Simulator ; for equivalent devices, follow the manufacturer’s instructions.
“Disassembling the IBP module” on page 12 • For instructions on removing the lower housing assembly from the M Series CCT unit or on replacing the multi-parameter interface cable, see the Non-Invasive Blood Pressure insert (part number 9653-0455-01) of the M Series Service Manual.
/22' 5(6685( 5HSODFHPHQW 3DUWV You can order replacement parts through an authorized ZOLL distributor or directly from ZOLL Medical Corporation. Pricing information is available from ZOLL Medical Corporation’s Technical Service Department. Note: ZOLL reserves the right to substitute different parts to reflect modifications and improvements in ZOLL M Series circuitry and design.
Page 25
7URXEOHVKRRWLQJ 7URXEOHVKRRWLQJ The M Series CCT unit displays certain error messages to alert the user to problems detected in IBP operation. The following table lists IBP error messages and their associated corrective action. For further assistance, contact ZOLL Technical Service at 1-800-348-9011. International customers should call the nearest authorized ZOLL Medical Corporation distributor.