Appendix C Messages - ZOLL M2 Operator's Manual

Table of Contents

Advertisement

This appendix lists the patient (physiological) alarms and equipment (technical) alarms that you
may see displayed on your ZOLL M2 monitor/defibrillator.
Alarm Message
4:1 BUTTON STUCK
30 J SELF-TEST BUTTON
STUCK
AED FUNCTIONS
DISABLED
ALARM BUTTON STUCK
ANALYZE BUTTON STUCK Equipment
ATTACH PADS
BATTERY CALIBRATION
REQUIRED
9650-000860-01 Rev. C
Alarm Type
Priority
Equipment
Medium
Equipment
High
Equipment
Medium
Equipment
Medium (Monitor/Pacer
mode;
High (Defib mode)
Medium (if "Restart analysis
after CPR" in Supervisor
menu is disabled);
High (if "Restart analysis
after CPR" in Supervisor
menu is enabled).
Equipment
Medium
Equipment
Low
ZOLL M2 Operator's Guide
Appendix C
Messages
Cause
Front panel quick access key self-test
failed.
Front panel quick access key self-test
failed.
Pads ECG streaming communication
test failed.
Front panel button self-test failed.
Front panel button self-test failed.
Pads are not connected to patient
(when pads is selected as primary
ECG waveform in Monitor mode or
Manual Defib mode).
The battery has reached a state that
requires battery calibration.
C-1