CLEANING, DISINFECTION, AND STERILIZATION

Cleaning
Clean the instrument as necessary by wiping the housing surfaces with a soft dry cloth or a soft
cloth that is lightly dampened with soapy water, 10% Clorox™/water solution, or 70% Isopropyl
alcohol. Clean the Patient Eyepiece window and Operator LCD Display window with a soft, lint
free cloth lightly dampened with commercially available window cleaners (do not use soap) or
70% Isopropyl alcohol.
Disinfection
Disinfect the patient contact surfaces (the Forehead Rest and Patient Response Button) as
necessary by wiping the surfaces with a soft cloth that has been lightly dampened with 10%
Clorox™/water solution or 70% Isopropyl alcohol. Allow the surface to dry thoroughly before
patient contact.
Note: Do not allow cleaning or disinfection solutions or other liquids to seep into the seams in
the housings or along the LCD display. Do not spray cleaning or disinfection solutions or other
liquids directly onto the instrument. Damage to internal components may occur and are not
covered by the instrument warranty.
Sterilization
Do not sterilize the instrument or any of its components
.
36
Humphrey Matrix Visual Field Instrument
Rev. D 08/21/06 PN 701692