Care Of The Unit; Cleaning Mechanical Surfaces; Sterilization - Zeiss S88 Instructions For Use Manual

Floor stand with lifting column
Table of Contents

Advertisement

156

Care of the unit

Cleaning mechanical surfaces

Sterilization

G-30-1426-en
Maintenance / Further information
Warning!
If possible, the systems and accessories should be cleaned immediately
after use. Contaminations should not be allowed to dry on the objects, as
this would make cleaning and disinfecting more difficult.
If possible, machines should be used for disinfecting and cleaning. For
details, please also see the relevant notes on sterilization equipment.
All mechanical surfaces of the equipment can be cleaned by wiping them
with a damp cloth. Do not use any aggressive or abrasive detergents.
Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled
water plus a dash of household dish-washing liquid.
The asepsis sets available from Carl Zeiss contain rubber caps, sleeves
and handgrips which can be sterilized in autoclaves. We recommend the
following program for sterilization:
Sterilization temperature: 134° C
Sterilization time:
Sterile single-use drapes are available to cover the system.
Note:
When draping the system, make sure there is enough slack in the drapes
to allow for movement of the microscope carrier and surgical microscope.
It is especially important that the drapes are completely loose around the
handgrips. The surgeon must be able to operate the controls through the
drape.
S88 Floor Stand with Lifting Column
10 minutes
Issue 9.0
Printed on 28. 01. 2009