Preparations for use

Connecting the system

CAUTION
R046
CAUTION
R057
NOTE
R017
Page 46
Increased leakage current!
There is a risk of user and patient sustaining an electric shock.
• Never touch the video plugs (connectors) while in contact with the pa-
tient.
• Do not connect any defective or unapproved accessories to the contacts
of connectors.
Risk of tripping over cables lying on the floor!
If cables are inappropriately installed on the floor, this increases the risk of
tripping and interferes with the flow of work. By tripping and falling down,
users may suffer tissue trauma.
• If possible, cables should only be laid outside of the area of movement.
Observe the rated voltage
Prior to leaving the factory, the device is adjusted to the rated voltage of the
respective country of destination. The set voltage is visible in a small
window (1) on the carrier arm. If the set voltage does not comply with the
line voltage available at the site of installation, this can cause damage to the
device.
• Check the voltage setting. The set voltage must comply with the line volt-
age available at the site of installation. If this is not the case, you must
not operate the system.
• If there is a variance, ZEISS Service will have to convert your device to the
line voltage available at your location.
Connecting the device cable
Make sure that the device has been switched off at the power switch before
plugging in or unplugging the device cable.
• Connect the device cable to the device by inserting connector (2) into
socket (4) on the device.
• Connect the device to line power by inserting the power plug (3) into a
socket of the examination room.
Only use a power outlet which is provided with a properly connected pro-
tective earth conductor.
KSK 150 FC
Version 7.1
G-30-1158-en