Connecting the suspension system

G-30-1426-en
Preparations for use
Check the voltage indicated at (3).
Caution:
The voltage of the stand is factory-set to the rated voltage used in the
country of destination. The rated voltage indicated at window (3) must cor-
respond to the rated voltage available at the site of installation. If this is
not the case, you must re-adjust the sliding switch using a suitable tool.
Note:
Insert or remove connectors (2) and (4) only if power switch (5) is off.
Plug the connector of the foot control panel or operating chair into con-
nector (2) of the stand.
Attach the microscope cable to the existing cable support in such a
way that it is not stretched or bent when the microscope is swiveled.
Secure the light guide in the cable support and insert the light into the
light guide mount on the microscope until stop. Make sure that the light
guide is not stretched or bent when the microscope is turned or
swiveled.
Connect external instruments with max. 24 V / 0.5 A to the remote control
socket (1); you must be able to switch these instruments on or off via an
AUX-signal to the foot control keys which can be configured as required.
Connect the stand to line power using the power cord intended for it. Only
use power outlets which are provided with a properly connected protec-
tive ground contact.
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009