14.9 CHANGING A DAMAGED PROBE
ROD
If the probe rod becomes damaged and will no
longer slide freely inside the probe bush, then it
must be replaced as shown below.
Loosen the grub screw securing the probe bush
to the main body of the probe then unscrew and
remove the probe bush.
Remove the probe rod and replace with a new
one.
Re-assemble the probe then calibrate as shown
Shown: Probe
in the calibration section.
The part number of the probe rod is 750/1042.
A probe repair kit is also available, part number
778/103.
80 PET Plus User Manual