Verifying the Analyzer

The verify process is used to correct for aging characteristics of the Sensor
assembly and potential fouling of the sensor by components of the sample
stream. By measuring the response of the sensor to an internally generated
"moisture standard, " any variations in sensor performance can be effectively
compensated for. During the verification operation, a new span factor is calcu-
lated for the sensor, and this value is then applied for all subsequent readings.

Changing Gases

If you are selecting a different gas, the analyzer will need time to purge the
internal volumes and to stabilize flow. To prevent false alarms from occurring
during this time, the display on the analyzer will read New Gas and the ana-
lyzer will automatically purge itself. This purge period lasts for forty minutes.
During this period, the Data Valid contact is opened and the flow alarms are
suppressed.
1-8 | 3050-RM Moisture Analyzer
AMETEK recommends performing a verification cycle monthly. How-
ever, no verification cycle should be performed during the startup of
the analyzer. It is recommended to allow the analyzer and internal
Moisture Generator to stabilize for a period of 24 hours after applying
power, before running a verification cycle.
You can abort this purge process by pressing the Verify key, scrolling
to Abort and then pressing Enter. If the process is aborted, normal
sampling will resume after the completion of the current measure-
ment cycle.
PN 305892901, Rev S